P-GM7 MicroStat Top-of-the-range sports analyzer, with pre-programmed assays for a detailed blood profile of the competitor. Assays include: Glucose, Lactate, Cholesterol, Alcohol, Urate, Pyruvate Glutamine, Ammonia |
![]() |
|
The Model P-GM7 is ideal for routine training monitoring, sports science teaching courses and research projects. A mains version, Model GM7, is also available. Analytical Principle: In oxidase enzyme reactions, the analyzer measures the rate of oxygen uptake and under appropriate controlled conditions, this is directly proportional to analyte concentration. Example:
Substrates for dehydrogenase enzymes are measured using two stage reactions in which the oxidation of NADH (produced or consumed) is monitored. Example:
Under the conditions of the assay, the rate of oxidation of excess NADH by peroxidase (POD) is inversely proportional to ammonia concentration.
Operation: For many assays, the simple injection of a sample is all that is needed to obtain a result and prepare the analyzer for the next analysis. Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds. The analyzer is first calibrated with a standard of the metabolite in question. Samples can be whole blood (e.g. key assays such as lactate, glucose, urea), plasma, serum or other biological fluids. Analyses are menu-driven via the 32-character display which guides the operator through the complete procedure. Subsidiary menus are reached via the keypad which enables the user to change analysis type, optimize operational modes, utilize special functions and perform statistical data analysis. The display also provides self-test diagnostics in relation to electrode status and fluid movements. Changeover between analytes is quickly and simply effected and all fluid pathways inside the analyzer can be rapidly sterilized without compromising performance. |
| Analytical Performance | |||
|
Glucose |
Lactate |
Alcohol |
| Precision (within run) |
Plasma; n=20,
CV=1.4% @ 10.0mmol/L Whole blood; n=40, CV=0.85% @ 12.0mmol/L |
Plasma; n=30,
CV=0.6% @ 10.0mmol/L Whole blood; n=22, CV=1.6% @ 5.0mmol/L |
Plasma; CV=1-2% @ 100md/dl Whole blood; CV=2-3% @ 85mg/dl |
| Accuracy - Recovery | Plasma; mean
100.9% (99.2-106.7%) range 1.5-30.0mmol/L Whole blood; mean 96.7% (89.1-100.2%) |
Plasma; mean
99.2% (96.0-104.0%) range 2.5-10.0mmol/L Whole blood; mean 96.7% (88.0-104.0%) |
Spiked serum; y (Analox) = 0.973 (true) + 2.1mg/dl r = 0.9929, n = 89 |
| Accuracy - Method Comparison, y = GM7 |
Plasma vs Hexokinase y = 0.96 (HK) - 2.5mg/dl r = 0.999; n = 156 (Plasma) |
Plasma vs Boehringer
UV, n = 20; r = 0.997 y = 0.967(UV) + 0.003 Whole blood vs YSI 23L, n = 56, r = 0.998 y = 1.017(YSI) - 0.087 |
Whole blood
vs GC, y (Analox) = 1.04(GC) + 1.3mg/dl r = 0.9905, n = 27 |
| Linearity | 0 - 30mmol/L
(540mg/dl) using 10µl sample 0 - 50mmol/L (900mg/dl) using 5µl sample |
0 - 10mmol/L
(90mg/dl) using 7µl sample 0 - 20mmol/L (180mg/dl) using 3.5µl sample |
0 - 200mg/dl
(43mmol/L) using 5µl sample 0 - 400mg/dl (86mmol/L) using 2.5µl sample |
| Instrument Specification: |
|
| Method Sensor Reaction temperature Display Printer |
Enzymatic oxygen-rate Clark-type amperometric oxygen electrode 30°C 32 character backlit LCD 16 column dot matrix, 1 line/sec |
| Statistical programs | a)Sequential, giving mean, S.D
and C.V. b) Selected results for 2 analytes, retains up to 75 entries |
| Interface Power Dimensions Weight |
Serial data port, optional Windows software
available 100-250V AC, 50-60Hz, 12-15V DC, 60VA 27cm (width) x 35cm (depth) x 25cm (height) 8.4Kg |
| Ordering Information: |
|
| P-GM7 MicroStat GM7 MicroStat GMRD-002A GMRD-011 GMRD-021 GMRD-084 |
Portable analyzer, rechargeable battery/mains Analyzer, mains version Glucose oxidase reagent, 360 analyzer cycles Glucose standard, 8mmol/L, 30ml (other values available) Uric acid reagent kit, 70 analyzer cycles incl. 0.6mmol/L standard Cholesterol reagent kit, 50 tests, incl. 5.2mmol/L standard |
| GMRD-090/093/092 | Lactate reagent for intact
whole blood - 100, 250 and 1000 analyzer cycles respectively, including 8.0mmol/L standard |
| GMRD-100/103/102 | Lactate reagent for lysed whole
blood - 100, 250 and 1000 analyzer cycles respectively, including 8.0mmol/L standard |
| GMRD-140 GMRD-150 GMRD-162 GMRD-050 GMRD-053/054 GMRD-055/056 |
Pyruvate reagent kit, 20-40
tests, incl. standard Glutamine reagent kit, 50 tests, incl. standard Ammonia reagent kit, 20-40 tests, incl. standard Record cards for collection capillaries, pack of 40 Blood collection capillaries, packs of 500 and 1000 respectively Closures for collection capillaries, packs of 100 and 500 respectively |
Specialized blood collection systems and accessories are available – please contact Analox for details |
|