The integral rechargeable battery allows the analyzer to be used at track or poolside and the optional all-weather carrying case enables the P-GL5 to be operated in virtually any environment. A mains version, the Model GL5, is also available. Analytical Principle: In oxidase enzyme reactions, the analyzer measures the rate of oxygen uptake and under appropriate controlled conditions, this is directly proportional to the concentration of the analyte. Example:
Operation: For each assay, the simple injection of a sample is all that is needed to obtain a result and prepare the analyzer for the next analysis. Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is then printed within 20 seconds. The analyzer is first calibrated with a standard of the metabolite in question. Samples can be whole blood (for key assays such as lactate and glucose), plasma, serum or other biological fluids Analyses are menu-driven via the 32-character display which guides the operator through the complete procedure. Subsidiary menus are reached via YES and NO buttons which enables the user to change analysis type, optimize operational modes, utilize special functions and perform statistical data analysis. The display also provides self-test diagnostics in relation to electrode status and fluid movements. Changeover between analytes is quickly and simply effected and all fluid pathways inside the analyzer can be rapidly sterilized without compromising performance. |
| Analytical Performance | |||
| Accuracy (Method Comparisons) | Linearity | Precision | |
| Glucose | y (Analox) =
0.98 (HK) - 2.5mg/dl n = 156, r = 0.998 |
0 -540mg/dl | CV = 1% @ 90mg/dl |
| Lactate | y (Analox) =
0.98 (YSI) + 0.8mg/dl n = 56, r = 0.998 |
0 - 90mg/dl | CV = 0.5 - 1.0% @ 50mg/dl |
| Alcohol | y (Analox) =
0.97 (True Val.) + 2.1mg/dl n = 89, r = 0.998 |
0 - 300mg/dl | CV = 2.5% @ 85mg/dl |
| Cholesterol | y (Analox) =1.06
(Man.Enz) - 7.7mg/dl n = 35, r = 0.998 |
0 - 400mg/dl | CV = 1% @ 200mg/dl |
| Urate | y (Analox) =
1.02 (AAII) +0.06mg/dl n = 53, r = 0.998 |
0 - 13.5mg/dl | CV = 2% @ 10mg/dl |
| Instrument
Specification: |
|
| Method Sensor Reaction Temperature Display Printer Statistics Interface |
Enzymatic oxygen-rate Clark-type amperometric oxygen electrode 30°C 32 character backlit LCD 16 column dot matrix, 1 line/sec. Sequential, giving mean, S.D and C.V. Serial data port, optional Windows software available |
| Power | 100-250VAC, 50-60Hz, 12-15VDC,
60VA, plus 5VA Integral rechargeable battery and charger unit |
| Dimensions Weight |
23cm(width) x 29cm(depth)
x 15cm(height) 6Kg |
| Ordering Information: |
|
| P-GL5 Portable GL5 Mains GMCC-098 GMRD-002A GMRD-011 GMRD-021 GMRD-084 |
Portable analyzer, rechargeable battery/mains Analyzer, mains version All-weather carrying case Glucose oxidase reagent, 360 analyzer cycles Glucose standard, 8mmol/L, 30ml (other values available) Uric acid reagent kit, 70 analyzer cycles incl. 0.6mmol/L standard Cholesterol reagent kit, 50 tests, incl. 5.2mmol/L standard |
| GMRD-090/093/092 | Lactate reagent for intact whole blood - 100,
250 and 1000 analyzer cycles respectively, including 8.0mmol/L standard |
| GMRD-100/103/102 | Lactate reagent for lysed whole blood - 100,
250 and 1000 analyzer cycles respectively, including 8.0mmol/L standard |
| GMRD-113 GMRD-110(100) GMRD-050 GMRD-053/056 GMRD-055/056 GMRD-074 |
Alcohol reagent kit, 70 analyzer cycles Alcohol standard, 100mg/dl, 4 x 1ml (other values available) Record cards for collection capillaries, pack of 40 Blood collection capillaries, pack of 100 and 500 respectively Closures for collection capillaries, pack of 500 and 1000 respectively Lactate quality control, 4 x 0.5ml |
Specialized blood collection systems and accessories are available – please contact Analox for details |
|